The Boaray5000 series ventilator has two models: Boaray5000C and Boaray5000D, the differences of configuration are shown as below:
Configuration/Model |
Boaray5000C |
Boaray5000D |
VCV mode |
Yes |
Yes |
PCV mode |
Yes |
Yes |
PSV mode |
Yes |
Yes |
SIMV mode |
Yes |
Yes |
SPONT mode |
Yes |
Yes |
DualPAP mode(High/Low pressure) |
Optional |
Optional |
PRVC mode |
Optional |
Optional |
NPPV mode |
Optional |
Optional |
Tidal volume |
50~1500 mL |
20~2000 mL |
Maintenance schedule
Maintenance |
|
|
|
|
Maintenance |
Frequency |
|
|
Every patient |
|
Respiratory valve, reusable breathing tube |
Every day |
|
The ventilator surface |
Every half year |
|
Fan filter, air water filter, battery |
Every year |
|
Oxygen sensor |
When cleaning installation |
and |
Check the parts good or not, if necessary change or repair it |
General |
|
|
|
Contents |
|
information |
|
|
Classification |
Ⅱb |
|
Electric shock defending type |
TYPE: Ⅰ, containing the internal power supply When you have doubt about the integrality of the external protective grounded or the protective grounded cable, the equipment should be replaced by the internal power supply (batteries) |
|
Electic shock |
|
|
|
TYPE B |
|
defending grade |
|
|
Suitabel range |
Pediatric(weight: 5 ~20kg) Adult(weight: 20 ~150kg) |
|
Power supply |
External power supply |
Power input: 100~240V,frequency 50/60Hz , Current input:1.5A (Max.) |
|
Backup batteries |
11.1VDC, 7800mAh |
|
Pressure supply range |
0.28~0.6MPa |
Gas supply |
|
The gas provided should not contain water, oil or impurity particle, |
|
Gas impurity |
and the content must be lower than the following standard. Air: H2O<7g/m3, Oil<0.5mg/m3 Oxygen: H2O<20mg/m3 |
Environment |
Working |
Temperature: +10 ℃~+50 ℃ Humdity: ≤95%(non-condensation) Atmospheric pressure: 50 kPa~106 kPa |
|
|
|
|
|
Temperature: -20℃~+55℃ |
|
Storage |
Humdity: ≤95%(non-condensation) |
|
|
Atmospheric pressure: 50 kPa~106 kPa |
The whole device |
Trolley |
Volume: 707×512×956(mm), Weight: 18.7kg(net weight) |
|
|
|
|
Host |
Volume: 489×443×541(mm), Weight: 23.2kg(net weight) |
Technical parameters
Description Parameter
Pneumatic electronically contronlled ventilator Category
Display |
15inch, TFT display, touch screen (optional) |
|
Battery |
Li batteries, power supply for about 120 minutes. |
|
Respiration mode |
VCV, PCV, SIMV+VCV, SIMV+PCV, SIMV+PRVC, PSV, NPPV(optional), PRVC(optional), DualPAP(optional) |
SPONT, |
Graphics display |
Waveform: prssure-time, flow-time,volume-time. Loop graph(optional):pressure-volume,flow-volume,flow-pressure. Others(optional): CO2, SPO2 |
|
Safe pressure in the airway system |
≤125cmH2O. |
|
Data communication interface |
RS232 serial port, communication interface, LVDS interface |
|
Sound pressure levels(Normal operation) |
≤ 65dB (A) |
|
Setting
parameters
Parameter |
Description |
|
Accurancy |
Tidal volume |
Range: 20~2000 mL (D series), Range: 50~1500 mL (C series) Increment: 20~100 Ml: 5 mL 100~1000 mL: 10 mL 1000~2000mL: 50 mL |
|
±20mL, or ±15% of the set value, whichever is greater |
Respiration frequency |
Range: in SIMV mode: 1~40 bpm In other modes: 4~100 bpm, increment: 1 bpm |
|
±2bpm or ±10% of the set value, whichever is greater |
Inspiration time |
Range: 0.1~10.0 s Increment: 0.1 s |
|
±0.1s or ±10% of the set value, whichever is greater |
Breathing hold time |
Range: 0~4 s, increment: 0.1 s |
|
|
Pressure trigger |
Range: -20cmH2O~-1 Increment: 1 cmH2O |
|
<-5 cmH2O: ±1 cmH2O, the other is ±20% |
Flow trigger |
Range: 0.5~20 L/min Increment: 0.5 L/min |
|
<-5 L/min : ±1 L/min, the other is ±20% |
PEEP |
Range: 0~50 cmH2O Increment: 1 cmH2O |
|
±2cmH2O or ±10% of |
Pressure support |
Range: 1~70 cmH2O Increment: 1 cmH2O Ppeak (pressure support) = pressure (Pressure support) + PEEP |
setting |
|
Range: 5~70 cmH2O |
the set value, |
Pressure control |
Increment: 1 cmH2O Ppeak (pressure control) = pressure setting |
whichever is greater |
|
(pressure control) + PEEP |
|
High pressure level(optional) |
Range: 5~70 cmH2O Increment: 1 cmH2O Not suitable for C series, and optional for D series |
|
Low pressure level(optional) |
Range: 0~50 cmH2O Increment: 1 cmH2O Not suitable for C series, and optional for D series |
|
High pressure support(optional) |
Range: 1~70 cmH2O Increment: 1 cmH2O |
|
Low pressure support(optional) |
Range: 1~70 cmH2O Increment: 1 cmH2O |
|
Oxygen concentration |
Range: 21~100% Increment: 1% |
±6% of the set value |
Apnea time |
Range: 10~60 sec Increment: 5 sec |
±1 s or ±10% of the set value |
Respiration sensitivity |
Range: 10~60% Increment: 10% |
|
Prate |
Range: 1~5 |
|
|
Increment: 1 |
|
Monitoring
parameters
Parameter |
Description |
Inspiration tidal volume |
Range: 0~3000 mL, Resolution: 1 mL |
Expiration tidal volume |
Range: 0~3000 mL, Resolution: 1 mL |
Minute ventilation volume |
Range: 0~99 L, Resolution: 0.1 L |
Spontaneous minute ventilation |
Range: 0~99 L, Resolution: 0.1 L |
volume |
|
Total respiration frequency |
Range: 0~100bpm, Resolution: 1 bpm |
Spontaneous respiration requency |
Range: 0~99 bpm, Resolution: 1 bpm |
Inspiration and expiration rate |
Range: 4:1~1:9, Resolution: 0.1 |
PEEP |
Range: 0~50 cmH2O, Resolution: 1 cmH2O |
Airway peek pressure |
Range: 0~100 cmH2O, Resolution: 1 cmH2O |
Average airway pressure |
Range: 0~100 cmH2O, Resolution: 1 cmH2O |
Inspiration platform pressure |
Range: 0~100 cmH2O, Resolution: 1 cmH2O |
Minimum airway pressure |
Range: 0~100 cmH2O, Resolution: 1 cmH2O |
Oxygen concentration |
Range: 21~100%, Resolution: 1% |
Airway resistance |
Range: 0~200 cmH2O/(L/s), Resolution: 1 cmH2O/(L/s) |
Compliance |
Range: 0~200 mL/cmH2O, Resolution: 1 mL/cmH2O |
Rapid shallow breathing |
Range: 0.0-1000.0 bpm/L, Resolution: 1 bpm/L |
Occlusion pressure P0.1 |
Range: 30>=P0.1>= 0, Resolution: 0.1 cmH2O |
Negative inspiration force |
Range: -40~0cmH2O, Resolution: 0.1 cmH2O |
AutoPEEP |
Range: 0~100cmH2O, Resolution: 0.1 cmH2O |
PR |
Range: 30~250 bpm, Resolution: 1bpm |
EtCO2 |
Range: 0~99 mmHg, Resolution: 1mmHg |
InCO2 |
Range: 0~99 mmHg, Resolution: 1mmHg |
Main alarm
parameters
Alarm content |
|
|
Range |
Tidal volume low |
High |
100~2000 mL, OFF |
|
Low |
OFF, 20~2000 mL |
|
|
Minute volume |
High |
0~40 L |
|
Low |
OFF, 0~39 L |
|
|
O2 concentration |
High |
21~100% |
|
Low |
OFF, 21~99% |
|
|
Airway pressure |
High |
6~100 cmH2O |
|
Low |
0~99cmH2O |
|
|
Frequency |
High |
1~100bpm |
|
Low |
0~99 bpm |
|
|
EtCO2 (optional) |
High |
1~99 mmHg |
|
Low |
OFF, 0~98 mmHg |
|
|
SPO2 (optional) |
Low |
80~100% |
|
O2 supply pressure |
High |
110~750 KPa |
|
Low |
100~740 KPa |
|
|
Air supply pressure |
High |
110~750 KPa |
|
|
Low |
100~740 KPa |
|
PRVC airway pressure upper limits |
Plimit≤Ppeak+5cmH2O |
|
|
High Continues Pressure |
It will give an alarm when the airway pressure is (PEEP+15)cmH2O for 15 seconds |
continuous above |
|
Apnea |
Range:10~60 s, increment :5 s |
|
|
O2 supply pressure low |
The O2 supply pressue is less than 0.28 MPa |
|
|
Air supply pressure low |
The O2 supply pressue is less than 0.28 MPa |
|
|
AC failure |
The alarm time is more than 120s |
|
|
Low battery capacity |
The battery working time is 10 minutes after alarm |
|
|
Battery exhausted |
The battery working time is 5 minutes after alarm |
|
|
Alarm silence count down |
≤100 s |
|
|
Alarm sound |
>65dB |
|
CO2module specification(optional)
Parameter |
Specification |
|
Measuring mode |
Mainstream |
|
Operation temperature |
0-40℃/ 32-104 ℉ |
|
Storage and transportation temperature |
-40-75℃/ -40-167 ℉ |
|
Operating humidity |
< 40 hPa H2O (non-condensing) (95 %RH at 30 °C) |
|
Storage and transportation humidity |
5-100 % RH, condensing |
|
Operation atmospheric pressure |
525-1200 hPa (525hPa corresponding to an altitude of 4572 m/15000 feet) |
|
Storage and transportation pressure |
500 to1200 hPa |
|
Total system response time |
<1 second |
|
Accuracy specifications(during |
Range |
Accuracy |
standard conditions) |
0 to 114 mmHg |
±(0.2 mmHg + 2% of reading) |